Qualmark Typhoon 2.0 HALT Chamber (LN2)

The Qualmark Typhoon 2.0, with its 24” x 24” vibration table, is perfect for performing Highly Accelerated Life Testing (HALT) on small products, or where limited lab space is available. This compact system is built using Qualmark’s Typhoon system technology which delivers impressive thermal performance and six-degree-of-freedom repetitive-shock vibration, and is highly efficient in LN2 usage and power consumption. The Qualmark Typhoon 2.0 is a practical addition to any company’s product reliability program.

The Qualmark Typhoon 2.0 is the perfect size for HALT testing small to medium electronics products. Whilst it’s size is small it’s impact on development is massive, you’ll save time, energy and frustration with the 2.0’s ability to quickly point the finger at weak spots in your designs.

 

6DOF repetitive-shock table

  • Simultaneous excitation at multiple frequencies for testing complex assemblies

  • Patented pneumatic hammer/table design is lubrication-free and valveless

  • Quiet operation without external cooling requirements

  • The table is field-adjustable for resetting to original PSD levels, ensuring consistent performance. (Adjustment included in our preventative maintenance agreement.)

 

Chamber Design

  • Stainless steel interior, painted exterior, with two full opening doors (one door on T1.5/2.0 models)

  • High-velocity airflow for improved product temperature change rates.

  • Two-position table for flexibility to test taller samples, improved ergonomics, and less LN2/electricity usage for smaller test items. (except T1.5/2.0)

  • Advanced chamber insulation combined with our low-noise hammers results in very quiet operation, even at maximum vibration levels.

  • Adjustable feet for leveling unit (stand options for T2.5/3.0 models)

 

Control System

  • Microsoft Windows system with Labview based operation and PLC-based controls

  • Programming, operation, and datalogging for vibration/thermal profiles.

  • One accelerometer for table control (three additional inputs available)

  • Two thermocouples: one for product, one for air (four additional inputs available)

  • Eight event relays for external controls

  • Optional Q-Link or Qualview software tools for ATE integration and other remote connections.

  • Optional integrated spectrum analyzers and data acquisition systems.

 

Liquid Nitrogen Vacuum Jacketed Piping

  • We work with your facilities teams to specify, design and supply the pipework for your HALT chamber.

  • Keep-full devices are essential to ensure optimum control of the chamber - preventing gas from creating PID disturbances.

  • Our pipe is made in the USA by Midwest Cryogenics and is simple to install and maintain.

  • Exhausts, airlines, and all other chamber utility requirements can be supplied.

  • External control and safety valves can be fitted to shut off supply in the event of a leak.

 

Safety Systems

  • We understand the risks of Liquid Nitrogen and have decades of experience planning installations and the safety mitigations that go with them.

  • Control panels inside the room are tied to gas detectors which can automatically shut the chamber supply line down at the source.

  • Control valves can be placed at key points along the route to ensure sections can be isolated through key facilities.

  • Additionally the control panels can be tied to facility supply and exhaust systems to increase air flow before critical alarms kick in and take the building down.

 

HALT Chamber Specifications (Qualmark Typhoon 2.0)

Qualmark HALT T2.0
Interior Volume 8.2 cu. ft.
233.0 Liters
Interior Dimensions
(W x D x H)
27" x 27" x 19.5"
685 x 685 x 495 mm
Exterior Dimensions
(W x D x H)
38.8" x 47" x 80.9"
985 x 1193 x 2054 mm
Temperature Range -100 to 200°C
-148 to 392°F
Average Change Rate
(lower with 50Hz power)
70°C/min
-65 to 85°C  
Repetitive Shock Vibration Acceleration: 5 - 75 gRMS (bare table)
Power 208V/230V 70A max (400V 3ph 50Hz)
Maximum Sound Level
(1 meter away)
73 dBA dBA
Notes Table size: 24" x 24" (610 x 610mm);
5 actuators

Qualmark T2.0 Datasheet Download

 