Thermal Shock Chamber (Large Capacity)
Large format thermal shock chambers allow for testing increasingly larger samples, quantities and samples that produce high heat-load.
Today, the automotive industry accelerates electronic multi-functions. According to this trend, safety standards such as ISO26262 and IEC61508 require high environmental stress that is generated by thermal shock chambers in order to bake in greater reliability to in-vehicle equipment.
Chambers are available in 1,100-liter, 1,650-liter, and 2,200-liter models. Unlike conventional chambers, this enables thermal shock testing of large automobile parts (e.g. Li-ion battery modules, electric turbochargers, intercoolers, radiators, and inverters for vehicles), large flat panel displays, and relatively large products that could not be previously tested due to their size or weight. This is ideal for customers who need to quickly test large amounts of specimens, such as for quality inspection during the production process, or research and development projects for the defence industry.
Model Summary
Model | TSA-1100S-W | TSA-1650S-W | TSA-2200S-W |
---|---|---|---|
Method | Two-zone by means of damper switching (three-zone optional) | ||
High-temp. exposure range | +60°C to +150°C | ||
Low-temp. exposure range | -50°C to -10°C | ||
Hot chamber max pre-heat setting temperature | +180°C | ||
Hot chamber temperature heat-up time | Within 40 minutes from ambient temperature to +180°C | ||
Cold chamber min pre-cool setting temperature | -65°C | ||
Cold chamber temperature pull-down time | Within 150 minutes from ambient temperature to -65°C | ||
Temperature recovery time | Two-zone | ||
High-temperature exposure: +85°C, 30 min. | |||
Low-temperature exposure: -40°C, 30 min. | |||
Power supply voltage: 200 V AC | |||
Sensor position: upstream | |||
Recovery rate: within 10 minutes | |||
Specimen (iron kg) | 50kg | 100kg | 100kg |
Test area | W1000mm | W1500mm | W2000mm |
H1100mm | H1100mm | H1100mm | |
D1000mm | D1000mm | D1000mm | |
Outside dimensions | W2120mm | W2620mm | W3120mm |
H1990mm | H1990mm | H1990mm | |
D2883mm | D2883mm | D2972mm | |
Weight | Approx. 3,400kg, | Approx. 3700kg | Approx. 3,900kg |
Power supply voltage | 200 V AC, 3 ohm, 3W, 50/60 Hz (Can support any power supply voltage as an option) | ||
Maximum current value | 125 | 180 | 198 |
Cooling water volume (25°C water) L/hr | 1540 | 2400 | 2400 |
Cooling water volume (32°C water) L/hr | 2800 | 4440 | 4440 |
Model | TSA-1100H-W | TSA-1650H-W | TSA-2200H-W |
---|---|---|---|
Method | Two-zone by means of damper switching (three-zone optional) | ||
High-temp. exposure range | +60°C to +180°C | ||
Low-temp. exposure range | -60°C to -10°C | ||
Hot chamber max pre-heat setting temperature | +200°C | ||
Hot chamber temperature heat-up time | Within 30 minutes from ambient temperature to +180°C | ||
Cold chamber min pre-cool setting temperature | -75°C | ||
Cold chamber temperature pull-down time | Within 150 minutes from ambient temperature to -75°C | Within 80 minutes from ambient temperature to -75°C | Within 90 minutes from ambient temperature to -75°C |
Temperature recovery time | Two-zone | ||
High-temperature exposure: +150°C, 30 min. | |||
Low-temperature exposure: -50°C, 30 min. | |||
Power supply voltage: 200 V AC, 3 ohm, 3W, 50/60 Hz (Supporting any power supply voltage as an option) | |||
Sensor position: upstream | |||
Recovery rate: within 10 minutes | |||
Specimen (iron kg) | 50kg | 100kg | 100kg |
Test area | W1000mm | W1500mm | W2000mm |
H1100mm | H1100mm | H1100mm | |
D1000mm | D1000mm | D1000mm | |
Outside dimensions | W2120mm | W2620mm | W3120mm |
H1990mm | H1990mm | H1990mm | |
D2883mm | D2883mm | D2972mm | |
Weight | Approx. 3,500kg, | Approx. 4,300kg | Approx. 4,300kg |
Power supply voltage | 200 V AC, 3 ohm, 3W, 50/60 Hz (Can support any power supply voltage as an option) | ||
Maximum current value | 172 | 270 | 310 |
Cooling water volume (25°C water) L/hr | 2400 | 4800 | 4800 |
Cooling water volume (32°C water) L/hr | 4440 | 8880 | 8880 |